BATTERY AND CHARGER TRANSPORTATION AND STORAGE CONDITION
Recommended Storage and Transportation Conditions 0°F (-18°C) to 113° F (45°C), up to 80% RH
CAUTION For optimal Battery life the Battery packs should be stored in an environment with low humidity, free from corrosive gas at a
recommended temperature <70°F (21°C). Extended exposure to temperatures above 113°F (45°C) could degrade Battery life
and performance.
NOTE Transport or store the Batteries and Chargers in an environment that limits exposure to dust, moisture and temperature extremes.
Electromagnetic Compatibility (EMC) GUIDANCE TABLES
TotalShield II
WARNING The
Surgical Helmets are suitable for use in Professional Heathcare Facility Environments except for near active HF
SURGICAL EQUIPMENT and the RF shielded room of a ME SYSTEM for magnetic resonance imaging.
TotalShield II
WARNING Use of the
improper operation.
WARNING Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no
closer than 30 cm (12 inches) to any part of
could result.
WARNING Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this equipment could
result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation.
The following tables provide guidance on electromagnetic compatibility for the
EMC GUIDANCE AND DECLARATION - EM EMISSIONS
Guidance and manufacturer's declaration – electromagnetic emissions
The
TotalShield II
Surgical Helmets are intended for use in the electromagnetic environments specified below. The customer or the user should assure that they
are used in such an environment. The EMISSIONS characteristics of this equipment make it suitable for use in industrial areas and hospitals (CISPR 11 class A).
If it is used in a residential environment (for which CISPR 11 class B is normally required) this equipment might not offer adequate protection to radio-frequency
communication services. The user might need to take mitigation measures, such as relocating or reorienting the equipment.
EMISSIONS TEST
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/Flicker emissions
IEC 61000-3-3
Surgical Helmets adjacent to or stacked with other equipment should be avoided because it could result in
TotalShield II
Surgical Helmets. Otherwise, degradation of the performance of this equipment
COMPLIANCE
ELECTROMAGNETIC ENVIRONMENT - GUIDANCE
The
Group 1
functionTherefore, their RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
Class A
The
N/A
domestic and those directly connected to the public low-voltage Power Supply network
that supplies buildings used for domestic purposes.
N/A
TotalShield II
Surgical Helmets
TotalShield II
Surgical Helmets use RF energy only for their internal
TotalShield II
Surgical Helmets are suitable for use in all establishments other than
7