Operating Design
4.8.8
Parameters in the "DEVICE/LEVELING" Menu
Parameter
Settings
LEVEL
OFF
NOTE*
ERR.MSG.
START
KEY
AUTO*
* Factory setting
4.8.9
Parameters in the "DEVICE/MOTION.S." Menu
Parameter
Settings
CONTROL
ON*
OFF
NUM.GEST.
2 GESTUR.*
4 GESTUR.
LEFT.CMD.;
RIGHT.CMD.;
UP.CMD.;
DOWN.CMD.
PRINT*
ZERO
TARE
IONIZER
D.SHIELD
OPEN LFT.
OPEN RT.
CLOSE DS.
NONE
* Factory setting
36
Operating Instructions Cubis
Explanation
The device does not display messages on the status of the level.
The [LEVELING] display flashes if the device needs to be leveled.
The [LEVEL.ERR.] message appears in the operating display if the device has
not been leveled.
Sets the trigger for leveling to "key": Leveling must be started manually by
pressing the [Leveling] button.
Sets the trigger for leveling to "automatic": The device automatically performs
a leveling function using the integrated inclination sensor before every
internal calibration/adjustment.
Explanation
Activates the motion sensor.
Deactivates the motion sensor.
Define the number of possible gestures.
Define one of the following functions for the gesture:
— Left
— Right
— Up
— Down
Execute the [Print] key.
Execute the [Zero] key.
Execute the [Tare] key.
Execute the [Ionizer] button.
Execute the [Change] key.
Open the left door of the draft shield.
Open the right door of the draft shield.
Close all draft shield doors.
Assign no function.