1.2 Combination possibilities
This prosthetic component is part of the Ottobock modular system. As
a result of its design, it may be combined with other prosthetic com­
ponents of the modular system. You can find possible restrictions in
this section. Contact the manufacturer with questions.
The product may only be combined with Ottobock prosthetic
components that are approved for waterproof additional pros­
theses.
2 Application
2.1 Indications for use
The product is to be used solely for lower limb prosthetic fittings.
2.2 Area of application
Approved for a body weight of up to 150 kg.
The product may only be used for waterproof additional pros­
theses.
The set screws of stainless steel can optionally be replaced by titani­
um set screws, which increase the protection against corrosion.
4WR95=1
The product must be used in TF prostheses only.
2.3 Environmental conditions
Allowable environmental conditions
Temperature range for use: -10 °C to +60 °C
Storage and transport conditions: -10° C to +60° C, relative humid­
ity: no restrictions
Humidity: fresh water, salt water – cleaning and drying is required
after contact and use.
Allowable immersion depth: 1 m
Unallowable environmental conditions
Sauna, extreme watersports
Mechanical vibrations or impacts
Unallowable environmental conditions
Perspiration, urine, acids or bases
Dust, sand, highly hygroscopic particles (e.g. talcum)
2.4 Service Life
In general, all modular adapters are tested by the manufacturer for
3 million load cycles. Depending on the patient's level of activity, this
corresponds to a service life of 3 to 5 years.
3 Safety
3.1 Explanation of warning symbols
Warning regarding possible risks of accident or
CAUTION
injury.
Warning regarding possible technical damage.
NOTICE
3.2 General safety instructions
CAUTION
Exceeding the service life
Risk of injury due to change in or loss of functionality and damage to
the product
► Ensure that the approved service life is not exceeded (see sec­
tion "Service life").
CAUTION
Mechanical damage to the product
Risk of injury due to change in or loss of functionality
► Use caution when working with the product.
► If the product is damaged, check it for proper function and readi­
ness for use.
► In case of changes in or loss of functionality, do not continue
using the product (see "Signs of changes in or loss of functional­
ity during use" in this section).