8. Standards
he Compex is based directly on me-
T
dical technology.
o guarantee your safety, the design,
T
manufacturing and distribution of
Compex are in conformity with the
requirements of the European Direc-
tive 93/42/CEE.
he device is in conformity with the
T
standard for general safety rules for
electromedical devices IEC 60601-1,
the standard for electromagnetic
compatibility IEC 60601-1-2 and the
standard for special safety rules for
nerve and muscle stimulators IEC
60601-2-10.
urrent international standards (IEC
C
60601-2-10 AM1 2001) require that a
warning be given concerning the ap-
plication of electrodes to the thorax
(increased risk of cardiac fibrillation).
irective 2002/96/CEE Waste electri-
D
cal and electronic equipment
(WEEE).

9. Patent

nap electrode: patent pending.
S

10. Normalised symbols

mportant: Under some condi-
I
tions, the effective figure for the
stimulation pulses can exceed 10 mA
and 10 V. The information given in this
handbook must be strictly observed.
he Compex is a class II device
T
unit with its own internal electric
power, with type BF applied sections.
aste electrical and electronic
W
equipment (WEEE).
he "On/Off" switch is a multi-
T
function key:
Functions
On/Off (2 positions, stable)
Waiting or preparation
for part of the unit
Stop (switching off)
166
Symbol N°(according to CEI 878)

11. Technical information

General points
ower supply NIMH rechargeable
P
battery (NIMH) (4,8 V ª 1200 mA/h)
hargers: The only chargers used for
C
recharging the battery bear the follo-
wing information:
Type TR503-02-E-133A03
Input 90-264 VAC / 47-63Hz / 0.5A max.
Output 9V / 400mA / 6W
Type TR1509-06-E-133A03
Input 90-264 VAC / 47-63Hz / 0.5A max.
Output 9V / 1.4A / 15W
Type TR503-02-A-133A03
Input 90-264 VAC /47-63Hz / 0.5A max.
Output 9V / 400mA / 6W
Type TR1509-06-A-133A03
Input 90-264 VAC / 47-63Hz / 0.5A max.
Output 9V / 1.4A / 15W
Type TR503-02-U-133A03
Input 90-264 VAC / 47-63Hz / 0.5A max.
Output 9V / 400mA / 6W
Type TR1509-06-U-133A03
Input 90-264 VAC / 47-63Hz / 0.5A max.
Output 9V / 1.4A / 15W
Neurostimulation
ll electrical specifications are given
A
for a charge between 500 and 1000
ohms per channel.
Channels: four independent and individually
adjustable channels electrically insulated from
each other and from ground
Impulse shape: constant rectangular current
compensated to eliminate any direct current
component to avoid any residual skin polarisation
Maximum current of an impulse: 100 milliamperes
Intensity increase interval:
– Manual stimulation intensity adjustment: 0-
694 (energy units)
– Impulse duration: 0.5 mA
From 50 to 400 microseconds
Maximum quantity of electricity per impulse: 80
microcoulombs (2 x 40 µC compensated)
Typical rise time for an impulse
01-03
3 microseconds (between 20% and 80% of
maximum current)
01-06
Impulse frequency
01-10
from 1 to 150 Hertz