Mettler Toledo IND570 Manual De Instalación página 61

Ocultar thumbs Ver también para IND570:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 109
21. Hold the body with a 20 mm wrench and use another 20 mm wrench to tighten the outer
22. Push and slide the exterior braided armor up the cable so that its cut end is tight against the
23. Install the clamp disassembled in step 11. Tighten the screws evenly so that the clamps tighten
24. Twist the green and black POWERCELL cable wires together with the internal drain wire, and
25. Feed the cables, except for the outer drain wire, into the enclosure through the opening from
2-28
METTLER TOLEDO IND570 Terminal Installation Manual
sealing part of the gland to 40 lb/in (5 Nm).
cable gland. Tidy up any stray wires by cutting and removing them.
symmetrically on the exterior braided armor. Note that the outer drain wire should emerge
between the clamp and the cable gland.
Figure 2-49: Exterior braided armor Clamped in Place
crimp on one of the included ring terminals.
Figure 2-50: Green and Black Wires with Internal Drain Wire
which the cable gland was removed.
Figure 2-48: Body of Gland Installed
30205319
|
11
|
07/2019

Hide quick links:

Publicidad

Tabla de contenido
loading

Tabla de contenido