The 2 mV/3 mV jumper settings are described in Table 2-13.
2.7.

Scale Capacity and Increment Display Options

2.7.1.
Metrology Line
In order to meet metrological requirements from different regions of the world, a metrological line
that shows capacity and increment information or Max/Min/e information on the IND570 display
can be enabled within terminal configuration (Figure 2-123). When enabled, this line is always
shown except when in the setup mode. If multiple ranges are selected, this line will scroll through
each of the available capacities and ranges.
Refer to the Scale and Terminal sections in the User's Guide Chapter 3, Configuration, for details on
enabling the Metrology Line in IND570. In cases where the metrology line of data is not accepted
by an agency, it can be disabled and a label with the required information may be installed near
the display.
2-70
METTLER TOLEDO IND570 Terminal Installation Manual
When removing the W1 jumper for 3mV/V operation, reposition it on just one of the pins, as
shown in Figure 2-121 and at left in Figure 2-122. If this jumper is not correctly positioned, the
plastic cover over the Main board will not fit properly. If this occurs, do not force the plastic
cover down. Remove the cover and position the jumper as shown.
Figure 2-122: W1 Millivolt Jumper Positions
Table 2-13: Millivolt Jumper Description
Jumper
Setting
OPEN
W1
CLOSED
Description
Sets the connected scale's load cell setting to 3 mV/V
Sets the connected scale's load cell setting to 2 mV/V.
30205319
|
11
|
07/2019