Mettler Toledo IND570 Manual De Instalación página 69

Ocultar thumbs Ver también para IND570:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 109
Both of these steps must be correctly completed to ensure the correct function of the PowerMount
modules.
2.4.5.8.1.
Cable and Gland Preparation, Harsh Enclosure
Ground and shield terminations are a critical part of the PowerMount system's immunity to noise
and electrical surges. Prepare and install the terminal end of the PowerMount cable as follows:
2-36
METTLER TOLEDO IND570 Terminal Installation Manual
Internal connection and grounding of the cable.
Figure 2-66: IND570 Harsh Enclosure with Cable Gland for PowerMount Connection
Figure 2-67: Cable Preparation Measurements
Table 2-8: Cable Preparation Lengths – Harsh
Letter
A
Length of inner braided shield
B
Length of red, white and blue wires
C
Length of black wire and drain wire
D
Amount of insulation to strip from wires
Description
Length
32 mm (1.25")
254 mm (10")
32 mm (1.25")
5 mm (0.2")
30205319
|
11
|
07/2019

Hide quick links:

Publicidad

Tabla de contenido
loading

Tabla de contenido