8. Place the ring terminals of the provided black wire and the incoming black, green and drain
9. Connect the ends of the blue, white, black and red wires, to the POWERCELL connector in the
2-32
METTLER TOLEDO IND570 Terminal Installation Manual
Table 2-6: Maximum POWERCELL Cable Lengths
Total Cell-Cell Cable
(meters/feet)
130/426
wires on the grounding stud inside the enclosure, and secure them in place using the nut with
star washer.
Figure 2-58: Provided Black Grounding Wire with Ring Terminal
Figure 2-59: Wires Connected to Internal Grounding Point
Note: Inside the enclosure it is important to keep the POWERCELL wiring, and especially the
ground wires, as short as possible to preserve lightning protection.
positions shown.
Home Run Cable
Number of
(meters/feet)
Cells
300/984
< 14
30205319
|
11
|
07/2019