Installation of a PowerDeck Platform Using Original Factory Calibration Values
During the original factory configuration of a PowerDeck floor scale, a complete set of System
Configuration Data for the platform is stored in each of the four (4) POWERCELL load cells in the
platform. The System Configuration Data set includes:
The IND570 has the ability to read this stored System Configuration Data and use it to create a
usable, basic weighing system without applying test weights during initial installation. This IND570
feature, known as Restore Scale Data, is intended for use when a new PowerDeck floor scale and
new IND570 POWERCELL terminal are combined and will be commissioned on-site.
In order for the terminal to use the original System Configuration Data stored in the PowerDeck load
cells, the following conditions must be true:
2.4.5.11.
Initial Commissioning Procedure
The steps below represent a typical PowerDeck installation procedure where the terminal's Restore
Scale Data will use the original System Configuration Data from the load cells to create a basic
weighing system.
1. Physically level the PowerDeck platform with a bubble level.
2. Shim, if necessary, to eliminate platform "rocking".
30205319
|
11
|
07/2019
This feature is only available in POWERCELL terminals, and should only be used when
connected to a PowerDeck floor scale.
• Version
• Number of load cells
• Platform serial number
• Platform model number
• Calibration weight unit
The IND570 POWERCELL terminal must be connected to a PowerDeck platform
The terminal is configured for use with a PowerDeck floor scale. The setting at Scale > Type >
Application must be Floor, and the # of Load Cells must be 4
The platform Model and Serial Number fields in the terminal are blank
Platform Model and Serial number fields can be viewed by pressing the RECALL
INFORMATION
softkey on the terminal Home page and then the SYSTEM
INFORMATION RECALL
2-4 of the POWERCELL load cells in the platform have matching System Configuration Data
If only 2 of the POWERCELL load cells have matching data, the remaining 2 cells must
have independently unique System Configuration Data. They cannot be another matched
pair. In a situation where there are of 2 matching pairs of System Configuration Data, the
terminal will not know which data set is original to the platform and should be used. In this
situation, neither set of data will be used and the Restore Scale Data function cannot
process.
The terminal is not metrologically sealed
• Factory calibration GEO value
• Field calibration GEO value
• Primary weight unit
• Scale capacity
• Scale increment size
softkey on the Recall screen.
METTLER TOLEDO IND570 Terminal Installation Manual
• Initial zero counts
• Test load 1 weight
• Test load 1 counts
• Load cell shift values for all cells
• Load cell calibration zero counts
2-45