Halyard COOLIEF Instrucciones De Uso página 4

Ocultar thumbs Ver también para COOLIEF:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 13
Cleaning and Sterilization Instructions
The COOLIEF* Multi-Cooled RF Module is a non-body contact device and therefore
is NOT sterile and CANNOT be sterilized. If cleaning is necessary, the surface of the
Module can be cleaned with a damp low-lint cloth with non-abrasive detergent
dissolved in water. Dry the surface after wiping down. NEVER spray or pour
liquids directly on the COOLIEF* Multi-Cooled RF Module.
Troubleshooting
The following table is provided to assist the user in diagnosing potential
problems.
PROBLEM
COMMENTS
No temperature
In order to
measurement
measure
in treatment
temperature, all
mode
devices must be
in good working
OR
order.
Inaccurate,
erratic or
sluggish
temperature
reading in
treatment
mode
No impedance
In order to
measurement
measure
in ready state
impedance,
the entire
system must
be connected,
and all devices
must be in good
working order.
Probe
Each connector
Connector does
is designed to
not fit in Probe
connect in a
Plug-In
specific way for
safety reasons.
If the connector
"keys" are out
of line, the
connectors
won't fit
together.
RF Generator
Each connector
Connector
is designed to
does not fit in
connect in a
RF Generator
specific way for
Plug-in
safety reasons.
If the connector
"keys" are out
of line, the
connectors will
not fit together.
4
TROUBLESHOOTING
• Ensure that all equipment
required is correctly connected.
• Check for an error message on
the RF Generator.
• Visually inspect the probe(s)
and cable(s) for damage. Ensure
that devices are dry and at
room temperature. If problem
persists, discontinue use.
• Ensure that all equipment
required is correctly connected.
• Check for an error message on
the RF Generator.
• Visually inspect the probe(s)
and cable(s) for damage. Ensure
that devices are dry and at
room temperature. If problem
persists, discontinue use.
• Check that the connector keys
are lined up in the proper
orientation.
• Check that the connectors
being connected have the
same number of pins and that
the pins are not damaged.
The COOLIEF* Multi-Cooled RF
Module is only to be connected
to Cooled RF Probes (SIP/
THP/LUP).
• Ensure that the connectors are
clean and unobstructed.
• Check that the connector keys
are lined up in the proper
orientation.
• Check that the connectors
being connected have the
same number of pins and that
the pins are not damaged.
The COOLIEF*
Multi-Cooled RF
Module is only to be connected
to the RF Generator V4.0 or
higher, featuring a 19-pin
connector.
• Ensure that the connectors are
clean and unobstructed.
Customer Service and Product Return Information
If you have any problems with or questions about this HALYARD* Equipment,
contact our technical support personnel:
Halyard Health
5405 Windward Parkway
Alpharetta, GA 30004 USA
E-mail: PMPorders@hyh.com
1-844-425-9273 (1-844-HALYARD)
Notes
In order to return products under limited warranty you must have a return
authorization number before shipping the products back to Halyard Health.
Limited Warranty
Halyard Health warrants that these products are free from defects in original
workmanship and materials. If these products prove to be defective in original
workmanship or original materials, Halyard Health, in its absolute and sole
discretion, will replace or repair any such product, less charges for transportation
and labor costs incidental to inspection, removal or restocking of product.
This limited warranty applies only to original factory delivered products that have
been used for their normal and intended uses. Halyard Health's limited warranty
shall NOT apply to Halyard Health's products which have been repaired, altered
or modified in any way and shall NOT apply to Halyard Health's products which
have been improperly stored or improperly installed, operated or maintained
contrary to Halyard Health's Instructions. The warranty period for HALYARD* RF
Probes and RF Generator Connector Cables is 90 days from the date of purchase,
unless otherwise stated.
Disclaimer and Exclusion of Other Warranties
There are no warranties of any kind, which extend beyond the description of
the warranties as previously mentioned. Halyard Health disclaims and excludes
all warranties, whether expressed or implied, of merchantability or fitness for a
particular use of purpose.
Limitation of Liability for Damages
In any claim or lawsuit for damages arising from alleged breach of warranty,
breach of contract, negligence, product liability or any other legal or equitable
theory, the buyer specifically agrees that Halyard Health shall not be liable for
damages for loss of profits or claims of buyer's customers for any such damages.
Halyard Health's sole liability for damages shall be limited to the cost to buyer of
the specified goods sold by Halyard Health to buyer which give rise to the claim
for liability.
The buyer's use of this product shall be deemed acceptance of the terms and
conditions of these limited warranties, exclusions, disclaimers and limitations of
liability for money damages.

Publicidad

Tabla de contenido
loading

Productos relacionados para Halyard COOLIEF

Tabla de contenido