Section E: Confirming The Current Valve Setting; Section F: Reading The Valve Setting With The X-Ray Overlay Tool; Injection - DePuy Synthes Codman Neuro CERTAS Plus Manual Del Usario

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 140
• Mark the patient's skin in the center of the valve
• Make sure the black lines on the Locator Tool line up
If these remedial steps fail to rectify the problem, replace
the valve.
SECTION E: Confirming the Current
Valve Setting
The setting of an implanted valve can be determined
by using the CODMAN CERTAS Tool Kit, and following
SECTION C: Post-Implantation Adjustment Procedure,
Steps 1 through 6.
An alternate method is to x-ray the valve. A proper x-ray
is generated when the film is shot perpendicular to the
plane of the valve with the non-implanted side of the
patient's head resting on the plate. The film must be taken
in relation to the valve and not the patient's anatomy. See
Figure 14 for x-ray views of the valve at each setting.
When viewing the x-ray film or screen to confirm
the valve setting, use the X-Ray Overlay Tool (See
SECTION F: Reading the Valve Setting with the X-Ray
Overlay Tool, Steps 1–5).
SECTION F: Reading the Valve Setting with the
X-Ray Overlay Tool
See Figure 15.
Note: Position the X-Ray Overlay Tool flush against the
x-ray image.
1. Align RED centerline of valve on overlay with the
2. Ensure that the numbers on the overlay that depict
3. Align rotating construct (RC) center dot on overlay
4. Ensure RHS marker red line containing red dot is aligned
5. The valve setting is determined by identifying the

Injection

If hypodermic injection is required, inject with a 25-gauge
or smaller non-coring HUBER
reservoir only. The reservoir can be punctured up to
25 times with a 25-gauge or smaller non-coring HUBER
type needle.
206521-001-G-INS.indd 17
mechanism between the two black lines. Also mark
the location of the valve barbs/catheters on the
proximal and distal ends (See Figure 7). This will
facilitate the proper position and orientation of either
Locator Tool above the implanted valve.
with the mark for the center of the valve mechanism
and that the direction of flow arrow is aligned with the
marks for the proximal and distal barbs/catheters.
centerline of the valve x-ray under review. This
can be accomplished by aligning the proximal and
distal connectors of the x-ray image with those on
the overlay.
the performance settings are properly oriented for
viewing. In this orientation the right hand side (RHS)
marker red line extends to the right of the RED
centerline. This ensures proper overlay orientation.
with the center of the RC of the x-ray image.
with the RHS marker of the x-ray image (if present).
region of the overlay that contains the majority of
the image of the magnet that has the tantalum ball
adjacent to it.
®
type needle into the
17
21-05-2015 23:30:56

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Codman neuro certas

Tabla de contenido