Valve Flushing (Clearing Obstructions)

(Valves without SIPHONGUARD Device only)
CAUTION: While flushing may be used as a method
for determining patency, it is not recommended.
Use clinical judgment and imaging studies or
other techniques to confirm suspected cases of
shunt malfunction.
To flush the ventricular (proximal) catheter:
• Occlude the catheter distal to the reservoir with finger
pressure, then depress the reservoir.
• Failure of the reservoir to refill might indicate proximal
catheter blockage.
To flush the distal catheter:
• Occlude the catheter proximal to the reservoir with
finger pressure, then depress the reservoir.
• Failure of the reservoir to depress easily might indicate
distal catheter blockage.
Note: While proximal pressure is maintained, reservoir
will not refill with fluid.

CODMAN CERTAS Tool Kit Replacement

The CODMAN CERTAS Tool Kit cannot be repaired.
Contact your local Codman representative for
replacement of the CODMAN CERTAS Tool Kit.

Warranty

Codman Neuro warrants that this medical device is
free from defects in both materials and workmanship.
Any other express or implied warranties, including
warranties of merchantability or fitness, are hereby
disclaimed. Suitability for use of this medical device
for any particular surgical procedure should be
determined by the user in conformance with the
manufacturer's instructions for use. There are no
warranties that extend beyond the description on
the face hereof.
® CODMAN, CERTAS, SIPHONGUARD and BACTISEAL are
registered trademarks of Codman Neuro, a division of DOI
® EXCITE is a registered trademark of General Electric Company
® PROLENE is a registered trademark of Ethicon, Inc.
® HUBER is a registered trademark of Becton, Dickinson, and Co.

Appendix A

Optional Manometer Testing

Although Codman does not recommend functional
testing, some surgeons might choose to do so. Before
testing, it is extremely important that a CODMAN CERTAS
Programmable Valve with or without SIPHONGUARD
Device be flushed of all air bubbles. Air bubbles within the
valve or SIPHONGUARD Device will produce inaccurate
manometer test results. The presence of air bubbles can
reduce the cross-sectional area of the flow path, increase
system resistance, and impede the flow of fluid through
the system during testing.

SIPHONGUARD Device Functional Testing

Equipment Required (use all sterile equipment, perform
testing under sterile conditions)
One sterile manometer, wide-bore (e.g. 3.5 mm),
graduated in mm (available in lengths from 38 to 60 cm)