Warning!
m
Shock hazard –
Strictly observe the following warnings.
Failure to do so endangers the lives of
the patient, the user and other persons
involved.
− Before use allow the ARTROMOT®-
E2/-E2 compact to reach room
temperature. If the device has been
transported at temperatures below
0 °C (32°F), leave it to dry at room
temperature for about 2 hours, until
any condensation has disappeared.
– The ARTROMOT®-E2/-E2 compact
must only be operated in dry rooms.
− When disconnecting the device from
the power line, remove the plug from
the wall outlet first, before discon-
necting the cable from the device.
− When connecting the device to
other equipment or when creating a
medical system, check that the sum
of leakage currents will not cause any
hazard. Please contact DJO, if you
have questions in this matter.
– Do not use multiple portable socket
outlets (MPSO) to connect the device
to the power line. The ARTROMOT®-E2/
-E2 compact must be connected to a
properly installed wall outlet with
a non-fused earthed wire. Before
connecting the power cord, it must be
completely unrolled and placed such
that it will not get caught in the mov-
ing parts of the device.
– Before cleaning and service interven-
tions, disconnect the device from
the power line by removing the
power cord from the wall outlet.
– Liquids must not be allowed to enter
the CPM device or the programming
unit. If liquids have entered into the
device, the ARTROMOT®-E2/-E2
compact must be checked by a service
technician, before it can be reused.
Warning!
m
Equipment malfunction –
− Magnetic and electrical fields are
capable of interfering with the proper
performance of the device. For this
reason make sure that all external
devices operated in the vicinity of the
CPM device comply with the relevant
EMC requirements. X-ray equip-
ment, MRI devices, radio systems, cell
phones, etc. are possible sources of
interference as they may emit higher
levels of electromagnetic radiation.
Keep the CPM device away from
these devices and verify its perfor-
mance before use.
− Refer repair and maintenance to
authorized persons.
− Route all cables such that they will
not get caught by the moving parts
during operation and that they do not
present a stumbling hazard.
− Inspect the ARTROMOT®-E2/-E2
compact for damage and loose con-
nections at least once a year.
Caution!
m
Preventing chafing and pressure sores –
If your patient is adipose, very tall or
very short, be sure to prevent chafing
and pressure sores.
Patient hazard, damage to the CPM
device –
The CPM device ARTROMOT®-E2 must
not be used to transport people.