DEVICE SPECIFICATIONS

Operating Characteristics

Power source: Lithium‐ion battery (rechargeable)
External power source: Powerbox EMX30
Input Power: 100 – 240 VAC, 47 – 63 Hz, 0.3 - 0.6 A
Output Power: 15 V, 2 A
Battery life*: 3 hours per charge; 5-years expected
lifetime
Dimensions: 257 mm (w) x 246 mm (h) x 22 mm (d)
Weight: 1125 g
Material:
Housing: Polycarbonate and ABS resin blend
Buttons: Silicone with polyurethane coating
Screen: Touch-screen, LCD display, 1280 x 800 pixels
CP Test Stimulation Output:
Maximum Amplitude: 12.5 mA
Frequency: 5 Hz or 14 Hz
Pulse width: 210 μs
*Note: Battery life may vary depending on frequency of use
110-0085 rev E.indd 9

Storage and Usage Environment

Usage environment
The following lists the appropriate temperature,
humidity, and pressure condition for use of the Axonics
CP:
Temperature: 5 °C to 35 °C
Humidity: 15% to 95%
Pressure: 70 kPa to 106 kPa
Shipping and Storage Environment
The following lists the appropriate temperature,
humidity, and pressure condition for shipping and
storage of the Axonics CP:
Temperature (short term: 3 days): ‐25
Temperature (long term): 20
Humidity (short term: 3 days): 15% to 95%
Humidity (long term): 30% to 85%
Pressure (short term: 3 days): 57 kPa to 106 kPa
Pressure (long term): 70 kPa to 106 kPa
If the CP is stored at temperatures outside of the
operating range, the CP should not be used until it has
come to the operating temperature range.
C to 70
C
o
o
C to 30
C
o
o
9 9