Axonics 2501 Manual Del Programador página 35

Ocultar thumbs Ver también para 2501:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 264
Tined Lead Test Stimulation
Recording the Motor and Sensory Responses
6
Motor and sensory responses associated with
stimulation at each electrode can be manually recorded.
Entering the motor and sensory responses is optional.
2
4
For example, if the patient is not awake during the
implant procedure, the sensory response can be left as
"None."
1
If saved, the response data and stimulation thresholds
5
8
are displayed when communicating with a Stimulator.
This data can be used to inform which stimulation
7
settings to use for therapy, and the CP will provide
3
recommended electrode configurations based on
this information. For example, if the stimulation with
a particular electrode elicited a "bad" sensation,
that electrode will not be included in recommended
electrode configurations.
Clear All Thresholds
Pressing the "Clear All" button -
( 7 ) ‐ will clear
established thresholds on all electrodes and allows this
process to be started over. This option may be useful if
the tined lead is repositioned, for example.
31 31
110-0085 rev E.indd 31
5/7/2020 11:53:28 AM