Controlling Stimulation Amplitude
20 20
110-0085 rev E.indd 20
The stimulation bar ( 1 ) controls the stimulation amplitude
during test stimulation for lead placement and Stimulator
2
programming. When available, the stimulation bar is on the right
side of the CP screen. The stimulation bar includes:
2 A digital display of the stimulation amplitude level.
1
When stimulation is OFF, this display shows in gray the
•
4
desired, or programmed, stimulation amplitude.
When stimulation is ON, this display shows in blue the
•
3
output, or delivered, stimulation amplitude.
Note: For extreme stimulation settings and impedance values (e.g., high
impedance values) the output stimulation amplitude may be less than the
programmed stimulation amplitude.
3 Up arrow and down arrow buttons to increase and decrease
stimulation amplitude incrementally.
By default, the "Variable" setting will allow for amplitude
•
changes in 0.05mA increments if the amplitude is
<1.30mA. At amplitudes above 1.30mA, the default
incremental change is 0.10mA.
On the CP Settings screen, the increment of amplitude
•
change can be changed from the "Variable" to a fixed
increment of 0.05mA or 0.10mA.
Incremental stimulation amplitude changes can be made
•
when stimulation is off or on.
4 A slider bar indicates the level of stimulation and can be
moved up or down to make large changes to the stimulation
amplitude.
Note: The slider bar can only be used to change stimulation amplitude when
stimulation is off.