Programming the Trial Stimulator
Delivering Test Stimulation
4
To deliver test stimulation, set the desired stimulation amplitude
on the stimulation bar ( 1 ) (see manual section Controlling
Stimulation Amplitude for more information). Then press the
stimulation button to deliver stimulation ( 2 ). The stimulation
button will light up when Stimulation is on.
1
When stimulation is on, the electrode configuration, ramp
settings, frequency, and pulse width cannot be changed.
Incremental changes can be made to the stimulation amplitude.
3
Save Therapy Settings
After therapy settings have been configured, press "Set
2
Program" ( 3 ) to save the settings to the Stimulator. A prompt
will appear for selecting the maximum amplitude the patient
can adjust their stimulation up to.
To exit, press "Patient Device" to return to the Patient Device
screen ( 4 ).
Note: If "Set Program" is not pressed prior to exiting, the Stimulator settings
will revert to last saved settings when the Programming screen is exited.
When the Programming screen is exited, a prompt will appear requesting
confirmation to exit without saving the therapy settings.
Note: The amplitude step size is fixed at 0.10mA.
53
110-0085 rev E.indd 53
5/7/2020 11:53:42 AM