Programming the Trial Stimulator
52
110-0085 rev E.indd 52
Adjusting parameters
1
Several adjustable stimulation parameters are shown in the
"Stimulation Parameters" box ( 5 ).
2
3
To adjust frequency and pulse width:
5
•
•
4
•
To adjust ramping:
•
•
Press the left and right arrows to the sides of the current value
to incrementally decrease and increase the parameter.
Frequency: the stimulation frequency can be adjusted from
2‐130 Hz. Adjustments are made in 1 Hz increments from 2‐50
Hz and 5 Hz increments from 50-130 Hz.
Pulse width can be set between 60 and 450 μs, adjustable in
10 μs increments.
Ramping settings are active when settings have been set to
the Stimulator.
Note: The ramping settings are not active during test stimulation. To test
ramping, save the stimulation settings to the Stimulator by pressing "Set
Program 1 or 2" then return to the Patient Device screen, and enable
stimulation.
Press the button for the parameter and a pop-up will appear
for adjusting the ramping setting.
Allows stimulation to ramp up and down between zero
•
amplitude and the targeted stimulation amplitude when
stimulation turns on and off.
Ramping time can be programmed at "Off", 1, 2, 4, 8, 15,
•
or 30 seconds.
5/7/2020 11:53:41 AM