Programming the Trial Stimulator
52
110-0085 rev E.indd 52
Adjusting parameters
1
Several adjustable stimulation parameters are shown in the
"Stimulation Parameters" box ( 5 ).
2
3
To adjust frequency and pulse width:
5
4
To adjust ramping:
Press the left and right arrows to the sides of the current value
to incrementally decrease and increase the parameter.
Frequency: the stimulation frequency can be adjusted from
2‐130 Hz. Adjustments are made in 1 Hz increments from 2‐50
Hz and 5 Hz increments from 50-130 Hz.
Pulse width can be set between 60 and 450 μs, adjustable in
10 μs increments.
Ramping settings are active when settings have been set to
the Stimulator.
Note: The ramping settings are not active during test stimulation. To test
ramping, save the stimulation settings to the Stimulator by pressing "Set
Program 1 or 2" then return to the Patient Device screen, and enable
stimulation.
Press the button for the parameter and a pop-up will appear
for adjusting the ramping setting.
Allows stimulation to ramp up and down between zero
amplitude and the targeted stimulation amplitude when
stimulation turns on and off.
Ramping time can be programmed at "Off", 1, 2, 4, 8, 15,
or 30 seconds.