Programming the Neurostimulator
Delivering test stimulation
4
To deliver test stimulation, set the desired stimulation amplitude
on the stimulation bar ( 1 ) (see manual section Controlling
Stimulation Amplitude for more information). Then press the
stimulation button to deliver stimulation ( 2 ). The stimulation
1
button will light up when Stimulation is on.
When stimulation is on, the electrode configuration, cycling
setting, ramp setting, frequency, and pulse width cannot be
changed. Incremental changes can be made to the stimulation
3
amplitude.
Save Therapy
2
After therapy settings have been configured, press "Set Device
Therapy" ( 3 ) to save the settings to the Neurostimulator.
To exit, press "Patient Device" to return to the Patient Device
screen ( 4 ).
Note: If "Set Device Therapy" is not pressed prior to exiting, the
Neurostimulator settings will revert to last saved settings when the
Programming screen is exited. When the Programming screen is exited,
a prompt will appear requesting confirmation to exit without saving the
therapy settings.