PNE Lead Test Stimulation
28 28
110-0085 rev E.indd 28
8
Note: The stimulation cable can be disconnected from the PNE lead without
causing stimulation to be turned off. This allows stimulation to be delivered
on demand by tapping the mini-clip portion of the cable against the
5
appropriate area of the lead.
Note: If test stimulation cannot be stopped using the stimulation button,
disconnect the cables from the CP.
6
1
9
4
7
2
3
To exit back to the Home screen, press "Exit" ( 9 ).
The area around the tip of the PNE lead graphic ( 5 ) will be
colored to reflect the proximity of the needle to the sacral
nerve. This information is used to evaluate the proximity of
the lead to the sacral nerve once the appropriate muscle
responses are observed (anal bellows and big toe flexion).
Gray
(unassigned): Stimulation level is zero.
Green
("good"): Stimulation level between 0.35 and 2.00
mA. Placement is acceptable.
Yellow
("OK"): Stimulation level between 0.05 and 0.30
mA or between 2.05 and 3.00 mA. Axonics recommends
adjusting the PNE lead placement.
Red
("not recommended"): Stimulation level between 3.00
and 12.50 mA. Axonics strongly recommends adjusting the
PNE lead placement.
If implanting a second PNE lead, press "Foramen Needle
Placement" to navigate back to the foramen needle test
stimulation screen to repeat the process of needle and lead
placement.
To connect to the Trial Stimulator, press "Connect to Patient
Device" ( 8 ) in the upper right.