DEVICE SPECIFICATIONS
Maintenance
At least once a year, the CP should be inspected for visible damage and should be charged and powered on to confirm
the Log-In screen is still accessible. Significant physical damage or an inability to power on the device should be
reported to the manufacturer and the device should not be used.
Handling and Disposal
Cleaning: The CP can be wiped with a cloth lightly moistened with water. No other cleaning agents should be used.
•
Replacement: If the CP is lost or not working, contact Axonics.
•
Disposal: Do not incinerate the CP as the battery may explode. If the CP is no longer needed, contact Axonics to
•
return the device.
Wireless Communication
Radiofrequency telemetry
•
Model: 2501
•
IC: 20225-C
•
Quality of Wireless Service:
•
This device operates in the 401‐406 MHz frequency and the maximum effective radiated power is below the
•
limit of 25 µW ERP/EIRP as specified in EU: EN ETSI 301-839 and EN ETSI 302-537 and USA: FCC 47 CFR Part 95;
Subpart I. The CP has to be within 1 meter from the Stimulator for successful communication.
Wireless Security:
•
Any CP can communicate with a Stimulator. Additional mechanisms exist to ensure the integrity of radio data.
•