allen A-71101-EU Manual Del Usuario página 15

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 321
If the Allen® Advance Table is to be used near high-frequency
surgical equipment, cardiac defibrillators, and cardiac defibrillator-
monitors, refer to the product manufacturer's instructions to make
sure the equipment will operate as intended.
Incorrect use or handling of the power cord may cause damage to
the power cord. If damage has occurred to the power cord or any of
its components, immediately remove the Allen® Advance Table
from service, and contact the applicable maintenance persons.
Do not expose the unit to excessive moisture.
Fluid spills on to the electronics can cause a hazard. If such a spill
occurs, unplug the unit and remove it from service. When fluid
spills occur outside that seen in normal use, immediately do as
follows:
a.
Unplug the unit from its power source.
b.
Remove the patient from the unit.
c.
Clean the fluid spill from the unit.
d.
Have maintenance examine the unit completely.
e.
Do not put the unit back into service until it is completely dry,
tested, and determined to be safe to operate.
Be aware of these pinch points on the Allen® Advance Table:
Between the caster and base skins
Between an H-bracket and the column
Between the head of the cross pin and the end cap during an
extreme tilt and Trendelenburg, or Reverse Trendelenburg
position.
Between the cross pin drop lock and the bowtie.
Between an H-bracket and the column during the removal of a
cross pin.
The unit is not intended for patient transport. However, if you must
use the unit to transport a patient, make sure that the unit is in its
lowest position and the patient is in a level position (not tilted).
Before you transport the unit, make sure of these:
The unit is in its lowest position.
The power cord is disconnected from the facility power outlet
and stored on the unit.
Accessories, supports, and any equipment that is to be
transported with the unit are within the perimeter of the unit
and secure on the unit.
Allen® Advance Table User Guide (D-720722 Ver D)
Safety warnings and cautions
15

Publicidad

Tabla de contenido

Solución de problemas

loading

Productos relacionados para allen A-71101-EU

Tabla de contenido