Specifications
Electromagnetic Immunity Guidance
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The A-71100/A-71101 are intended for use in the electromagnetic environment specified
below. The customer or the user of the A-71100/A-71101 should make sure it is used in such
an environment.
Immunity Test
Electrostatic
Discharge (ESD)
IEC 61000-4-2
Electrical Fast
Transient/Burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions, and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power
frequency(50/60
Hz) magnetic field
IEC 61000-4-8
Note: U
is the AC mains voltage prior to application of the test level.
T
58
IEC 60601 Test
Level
± 6 kV Contact
± 8 kV Air
± 2 kV for power
supply lines
± 1 kV for
input/output lines
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
< 5% U
T
(> 95% dip in U
)
T
for 0.5 cycles
40% U
T
(60% dip in U
) for
T
5 cycles
70% U
T
(30% dip in U
) for
T
25 cycles
< 5% U
T
(> 95% dip in U
)
T
for 5 seconds
3 A/m
Allen® Advance Table User Guide (D-720722 Ver D)
Compliance
Level
± 6 kV Contact
± 8 kV Air
± 2 kV for power
supply lines
± 1 kV for
input/output lines
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
< 5% U
T
(> 95% dip in U
)
T
for 0.5 cycles
40% U
T
(60% dip in U
) for
T
5 cycles
70% U
T
(30% dip in U
) for
T
25 cycles
< 5% U
T
(> 95% dip in U
)
T
for 5 seconds
3 A/m
Electromagnetic
Environment—
Guidance
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Mains power quality
should be that of a typical
commercial or hospital
environment.
Mains power quality
should be that of a typical
commercial or hospital
environment.
Mains power quality
should be of a typical
commercial or hospital
environment. If the user of
the A-71100/A-71101
requires continued
operation during power
mains interruption, it is
recommended to power the
equipment from an
uninterruptedly power
supply or a battery.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.