INSTRUCTIONS FOR USE
6.
Adjust the supports widthwise such that the inner edge of the support is
approximately aligned with the patient's acromion, then tighten the
adjustment knob at the top of the support.
7.
Place the table in Trendelenburg and allow the patient to settle into their final
position, as specified in the Hug-u-Vac Steep Trend instructions.
8.
Once the patient has settled, return the table to horizontal position and adjust
the supports to maintain the 1" gap specified in step 4., if required.
Device controls and indicators:
Left/Right Label
CE Label
IFU Matl Number: D-720674
Document Number: D-720674
Version: C
Serial Number
Part Number
Page 15
Usage Information Label
Issue Date: 13 MAR 2020
Ref Blank Template: 80025117 Ver. F