Integra CUSA Excel Manual Del Usuario página 4

Puntas estériles
Ocultar thumbs Ver también para CUSA Excel:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 21
EN
Types of Tips
Sterile tips are available at two frequencies: 23 kHz
and 36 kHz. Use the 23kHz sterile tips with the
C2600 23 kHz straight handpiece or C2601 23 kHz
angled handpiece. Use the 36kHz sterile tips with
the C2602 36kHz handpiece. Use the tables below
to select the correct tip for your handpiece:
23kHz Sterile Tips
Description
C4600S
MicroTip™
C4601S
Standard
Tip
C4602S
Straight
Extended
Standard
Tip
C4605S
MacroTip™
For use with C2600 23 kHz straight hand-
piece only:
C4603S
Curved
Extended
Standard
Tip
36kHz Sterile Tips
Description
C4606S
Precision-
Tip™
C4607S
Straight
Extended
PrecisionTip
C4608S
Curved
Extended
PrecisionTip
C4609S
MicroTip™
C4610S
Straight
Extended
MicroTip
C4611S
Curved
Extended
MicroTip
C4612S
Standard Tip
C4613S
Straight
Extended
Standard Tip
4
Sterilization Information
Supplied sterile. All items in this pack are for single
patient use only. Sterile tips may be resterilized
one time if unused. Steam sterilize only. Refer to
the CUSA Excel System User's Guide for the steam
Diameter
sterilization parameters.
1.57 mm
1.98 mm
For single patient use only. Do not reprocess or
1.98 mm
re-use.
Device(s) is (are) intended to be used for one
procedure only. If reprocessed or re-used this
may result in the infection of patient (or patient
specimen) through cross-contamination, as
2.64 mm
well as would incur the risk of modifying the
properties and performance of the device, and of
increasing the likelihood of complications and/or
undesirable effects. Once used, devices must be
disposed of in accordance with hospital policies.
1.98 mm
Duty Cycle
Under maximum loading conditions, the
CUSA Excel System is suitable for ultrasonics
activation times of 10 minutes on, 5 minutes off.
Assembly Options
Diameter
You have two options:
1.14 mm
1. NOT Changing Tips on the Sterile Field — If
1.14 mm
1.14 mm
2. Changing Tips on the Sterile Field-These
1.57 mm
1.57 mm
1.57 mm
Changing Tips on the Sterile Field
Assembly
1.98 mm
For this procedure, use the sterile, disposable
1.98 mm
torque wrench (C5601 for 23 kHz or C5602 for
36 kHz) and the sterilizable torque base (C5623 for
23 kHz or C5636 for 36 kHz).
C4614S
Curved
Extended
Standard Tip
Warning
assembling the tip to the handpiece in a
nonsterile area before sterilization, refer to the
following sections in the CUSA Excel System
User's Guide:
• Assembling the Handpiece in a Nonsterile
Area
• Completing Handpiece Setup in the Sterile
Field
instructions tell you how to assemble and
disassemble the handpiece as part of changing
tips on the sterile field.  For further information,
refer to section Assembling or Changing Tips on
the Sterile Field in the CUSA Excel System User's
Guide.
1.98 mm

Publicidad

Tabla de contenido
loading

Tabla de contenido