Priming The System - Integra CUSA Excel Manual Del Usuario

Puntas estériles
Ocultar thumbs Ver también para CUSA Excel:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 21
Testing the Handpiece
Once you have completed system setup, test the
handpiece to ensure that you have assembled
the tip correctly. Refer to the CUSA Excel System
User's Guide for instructions on how to test the
handpiece.

Priming the System

Prime the irrigation system as described in the
CUSA Excel System User's Guide. Repeat Prime
until you see irrigation fluid at the handpiece tip.
Disassembly
Use this procedure only if you are disassembling
the handpiece as part of changing tips on the
sterile field. If you are disassembling the handpiece
at the end of the surgical procedure, refer the
CUSA Excel System User's Guide.
1. Press the Status button to put the system in
Standby.
Disconnecting the Manifold Tubing:
1. Disconnect the irrigation tubing from the
handpiece flue.
2. Remove the flue.
3. Disconnect the manifold suction tubing and its
clip from the handpiece.
Removing the Nosecone, O-rings, and Tip:
1. Remove the nosecone:
a. Release the nosecone by twisting it counter
clockwise.
b. Pull the nosecone away from the handpiece.
Status
button
®
DO NOT DISCARD A STANDARD NOSECONE.
2. Remove the o-rings.
3. Locate the sterile torque base in which the
handpiece support color matches the color
on the handpiece connector (23 kHz — green;
36 kHz — blue). Put the handpiece in the
torque base so that the metal connecting body
fits snugly in the metal slot at the end of the
support.
4. Hold the handpiece securely in place in the
torque base.
Notice
To avoid product damage, NEVER hold the
handpiece in your hand while using the torque
wrench to tighten or loosen the tip.
5. Match the colored ring of the sterile torque
wrench with the handpiece connector color or
handpiece support color on the torque base.
Slide the color coded side of the wrench over
the tip until the hex in the wrench engages the
hex of the tip.
Blue ring on head of 36
kHz wrench
Green ring on head of
23 kHz wrench
EN
7

Publicidad

Tabla de contenido
loading

Tabla de contenido