As a second step, check the illumination system of the scope.
Reduced brightness can result from fiber damage:
• Check for fiber optic damage in the scope by holding the distal
end of the scope toward a low power light and observing the light
post on the hub. The center of the light post should appear clear or
white. Noticeable black spots indicate serious damage to the fiber
illumination bundle in the scope. This will affect light transmission
and the brightness of the image viewed on the monitor.
• Check the light cable for damaged fibers by holding one end
of the cable toward a low power light and observing the other
end. Broken fiber will appear as black spots in the light field. A
damaged light cable will affect its ability to transmit light and the
brightness of the image viewed on the monitor.
Storage
The MyoSure Rod Lens Hysteroscope and Removable Outflow
Channel should be stored either in their shipping box or in a
sterilization tray. In either case, proper care should be taken to
ensure that the hysteroscope and outflow channel are immobile to
prevent any damage.
Service - Accessories
The following are replacement/service parts for the MyoSure Rod
Lens Hysteroscopes and Removable Outflow Channels:
REF
Description
40-201
Replacement MyoSure Outflow
Channel
50-201XL
Replacement MyoSure XL Outflow
Channel
ASY-04996
Hysteroscope Light Source Adapters -
1 each: Wolf and Storz
40-902
MyoSure Single Use Seal Set - 10 per
box
40-904
MyoSure Hysterscope and Outflow
Channel Seal Cap
WARRANTY, SERVICE, AND REPAIR
WARRANTIES
Except as otherwise expressly stated in the Agreement: i) Equipment
manufactured by Hologic is warranted to the original Customer
to perform substantially in accordance with published product
specifications for one (1) year starting from the date of shipment,
or if Installation is required, from the date of Installation ("Warranty
Period"); ii) digital imaging mammography x-ray tubes are warranted
for twenty-four (24) months, during which the x-ray tubes are fully
warranted for the first twelve (12) months and are warranted on a
straight-line prorated basis during months 13-24; iii) replacement
parts and remanufactured items are warranted for the remainder of
the Warranty Period or ninety (90) days from shipment, whichever is
longer; iv) consumable Supplies are warranted to conform to published
specifications for a period ending on the expiration date shown
on their respective packages; v) licensed Software is warranted to
operate in accordance with published specifications; vi) Services are
warranted to be supplied in a workman-like manner; vii) non-Hologic
Manufactured Equipment is warranted through its manufacturer and
such manufacturer's warranties shall extend to Hologic's customers,
to the extent permitted by the manufacturer of such non-Hologic
Manufactured Equipment. Hologic does not warrant that use of
Products will be uninterrupted or error-free, or that Products will
operate with non-Hologic authorized third-party products.
These warranties do not apply to any item that is: (a) repaired,
moved, or altered other than by Hologic authorized service personnel;
(b) subjected to physical (including thermal or electrical) abuse,
stress, or misuse; (c) stored, maintained, or operated in any manner
inconsistent with applicable Hologic specifications or instructions,
including Customer's refusal to allow Hologic recommended Software
upgrades; or (d) designated as supplied subject to a non-Hologic
warranty or on a pre-release or "as-is" basis.
TECHNICAL SUPPORT AND PRODUCT RETURN INFORMATION
Contact Hologic Technical Support for proper disposal of any part
of the MyoSure Hysteroscopic Tissue Removal System or if the
MyoSure Hysteroscopic Tissue Removal System fails to operate
as intended. If product is to be returned to Hologic for any reason,
Technical Support will issue a Returned Materials Authorization
(RMA) number and biohazard kit if applicable. Return the MyoSure
Hysteroscopic Tissue Removal System according to the instructions
provided by Technical Support. Be sure to clean and sterilize the
product before returning it and include all accessories in the box
with the returned unit.
Return used or opened product according to the instructions
provided with the Hologic-supplied biohazard kit.
For More Information
For technical support or reorder information in the United States,
please contact:
Hologic, Inc.
250 Campus Drive
Marlborough, MA 01752 USA
Phone: 1.800.442.9892 (toll-free)
www.hologic.com
International customers, contact your distributor or local Hologic
Sales Representative:
European Representative
Hologic BVBA
Da Vincilaan 5
1930 Zaventem
Belgium
Phone: +32 2 711 46 80
Hologic, MyoSure, and associated logos are registered trademarks
of Hologic, Inc. and/or its subsidiaries in the United States and other
countries. All other trademarks, registered trademarks, and product
names are the property of their respective owners.
© 2014-2020 Hologic, Inc.
5
ENGLISH
AW-07813-001 Rev. 010

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Myosure 50-250xl

Tabla de contenido