Switching modes 12K50
When the battery is inserted, the current
switching mode is indicated by vibration
feedback signal.
Switching
Lock
mode
Press the switch ➜ release = lock
1
white
Press the switch ➜ release = unlock
Co-contraction = lock
2
red
Co-contraction = unlock
Press and hold down switch = elbow mode
3
green
Release the switch = Hand mode
Press the switch ➜ release= elbow mode
4
blue
Press the switch ➜ release = hand mode
Press the switch ➜ release= elbow mode
5
yellow
10 s no electrode signal = hand mode
or press the switch ➜ release = hand mode
Co-contraction = elbow mode
6
magenta
Co-contraction = hand mode
Co-contraction = elbow mode
7
orange
10 s no electrode signal = hand mode
or co-contraction = hand mode
Vibration feedback signal for successful switching over
between hand and elbow is activated in Programs 3 – 7.
1x vibration = hand mode (the hand is controlled by electrode signals)
2x vibration = elbow mode (the elbow is controlled by electrode signals)
Electrode OPEN = unlock
Electrode CLOSE = lock
Electrode OPEN = unlock
Electrode CLOSE = lock
Electrode OPEN = unlock
Electrode CLOSE = lock
Electrode OPEN = unlock
Electrode CLOSE = lock
Electrode OPEN = unlock
Electrode CLOSE = lock
Four Channel
Processor II
SLIP-STOP
NO
All switching modes
Only Program 1 (white)
Four-channel Control
NO
or
Program 9
Adjustment Cap
YES
YES
All switching modes
YES
YES
Only Program 1 (white)
Four-channel Control
or
Program 9
Adjustment Cap
YES
Hand
All versions
All versions with
2 electrodes
Not
recommended
for hands with
Digital or
Digital-Twin
Control System
49