11. Troubleshooting
THE UNIT DOES NOT OPERATE
PROPERLY
THE MONITOR DOES NOT OPERATE
PROPERLY
THE MICROMOTOR DOES NOT START
UP
THE PEDAL DOES NOT START UP THE
MOTOR
Ensure the cable is properly plugged into
the power socket and the network voltage
corresponds to the voltage indicated on the
power supply labelling.
Press Function
and Function
restore factory parameters.
If the monitor tends to become less bright,
it means ATR EndoPocket is distributing the
maximum power available. In this case, check
the integrity of the contra-angle.
Ensure the motor connector is properly housed
in the micromotor housing (red).
Check the functional integrity of the contra-
angle.
Remove the contra-angle and set maximum
speed, then start up the motor again.
Calibrate without the contra-angle, then connect
the contra-angle again and calibrate.
Try starting up the motor through the button on
the keyboard. In the event the motor starts up,
call the service centre to replace the pedal.