6.2.3.
Setting by tolerance values
13
12
11
10
1 Save value as a favourite.
2 The length entered in "Length setting"
3 Actual operating temperature
4 Internal value / external value selected in "Length
setting"
5 Time
6 Unit of measure
7 Move to the Max value
ü All accessory parts removed.
ü No objects or parts of the body between the ceramic gauge blocks.
ü "Length setting" is selected.
1. Use the numeric field to input the Nominal bore diameter d.
2. Select "ENTER" to move to the length that was input.
3. Confirm the value by pressing "Go".
»
The blocks will move to the value.
4. Select "TOL".
5. Select and input "ISO" or "Quality".
»
The values Max, Med and Min will be calculated.
6. Travel to the value by pressing "ENTER".
7. Confirm the value by pressing "Go".
»
The blocks will move to the value.
1 2 3 4
0012.000
Nominal bore diameter d
0012.000
mm
ISO fit
Quality
ISO
--.- °C
mm
Max
ENTER
----.---
mm
Med
ENTER
----.---
mm
Min
ENTER
----.---
mm
8 Move to the Med value
9 Move to the Min value
10 Calculation by Quality
11 Calculation by ISO
12 The Nominal bore diameter d corresponds to the
length
13 Back to "Length setting"
5
08:41
6
7
8
9