TROUBLE‐SHOOTING 
 
Anomaly 
 
The device does not 
switch on 
 
 
 
The device switches 
on but it does not 
produce water 
(pump stopped, no 
pressure to the 
pressure gauge)  
The device produces 
a small amount of 
water or produces no 
water at all (regular 
pressure to the 
pressure gauge) 
The device produces 
a small amount of 
water  (regular or 
high pressure to the 
pressure gauge) 
The device does not 
produce water  
(regular or high 
pressure to the 
pressure gauge) 
 
Biox does not switch 
on 
 
Possible cause
Power supply 
missing 
Plug detached
Fuse damaged
Machine 
damaged 
Tank full, 
Machine OK 
 
Infeed water 
missing  
Minimum 
pressure switch 
defective  
 
Discharge valve 
open  
Exhausted filters 
or membranes  
Exhausted filters 
or membranes 
Power supply 
missing 
Plug detached
Fuse damaged
Routine solution 
(Operator) 
Plug power supply
Attach the plug
Replace the fuse
‐‐‐
Open a utility and 
check for restarting  .
Check the water 
infeed valve and 
water presence in 
the waterworks   
‐‐‐
 
Close the discharge 
valve (light ON)   
‐‐‐ 
‐‐‐ 
Plug power supply
Attach the plug
‐‐‐
46
Required 
intervention 
(Skilled engineer)
‐‐‐ 
‐‐‐ 
‐‐‐ 
Repair the 
damage 
‐‐‐ 
‐‐‐ 
Check and replace 
the minimum 
pressure switch  
 
‐‐‐ 
Replace filters 
and/or 
membranes  
Replace filters 
and/or 
membranes 
‐‐‐ 
‐‐‐ 
Replace the fuse 