ENGLISH
TROUBLE‐SHOOTING
Anomaly
The device does not
switch on
The device switches
on but it does not
produce water
(pump stopped, no
pressure to the
pressure gauge)
The device produces
a small amount of
water or produces no
water at all (regular
pressure to the
pressure gauge)
The device produces
a small amount of
water (regular or
high pressure to the
pressure gauge)
The device does not
produce water
(regular or high
pressure to the
pressure gauge)
Biox does not switch
on
Possible cause
Power supply
missing
Plug detached
Fuse damaged
Machine
damaged
Tank full,
Machine OK
Infeed water
missing
Minimum
pressure switch
defective
Discharge valve
open
Exhausted filters
or membranes
Exhausted filters
or membranes
Power supply
missing
Plug detached
Fuse damaged
Routine solution
(Operator)
Plug power supply
Attach the plug
Replace the fuse
‐‐‐
Open a utility and
check for restarting .
Check the water
infeed valve and
water presence in
the waterworks
‐‐‐
Close the discharge
valve (light ON)
‐‐‐
‐‐‐
Plug power supply
Attach the plug
‐‐‐
46
Required
intervention
(Skilled engineer)
‐‐‐
‐‐‐
‐‐‐
Repair the
damage
‐‐‐
‐‐‐
Check and replace
the minimum
pressure switch
‐‐‐
Replace filters
and/or
membranes
Replace filters
and/or
membranes
‐‐‐
‐‐‐
Replace the fuse