Press EXIT to return to the upper level of the menu.
The entries are rejected.
4.8.2  E ntering IDs / PIN
The four softkeys and READ key are used for navigating on the keypad and entering the ID/PIN.
Example: WATER
1
If 1 is highlighted, press
Q is highlighted.
2
Press
W is highlighted.
3
Press READ to enter W.
4
Repositioning the highlighted bar to A, T, E and R, and press READ to enter each letter of sam-
ple ID in sequence as described in steps 1 - 3.
5
Reposition the highlighted bar to OK, and press READ to save the sample ID.
Instead of entering an ID with the alphanumeric keypad, one can also use a USB-keyboard or a
USB-barcode scanner. In case a character is entered or scanned that is not present on the instru-
ment keyboard it will display the entry as an underscore (_).
4.8.3  E diting values in a table
The meter has a feature, which allows the user to enter, edit or remove values in tables. (for exam-
ple, temperature and buffer values for a customized buffer group). This is accomplished by using
the softkeys on the display to navigate from cell to cell.
1
Press READ to start editing the cell in the table.
The softkeys on the display change.
2
Press
The softkeys change back to
3
Navigate to a cell and press Delete to remove a value.
4
To finish editing the table, navigate with the
5
Press READ to confirm the action and exit the menu.
4.9  C alibration
Calibration is only possible in the full-information screen. When starting a calibration by pressing
the CAL key while the instrument displays the close-up screen, it will automatically switch to the
full-information screen.
1
Place the electrode in a calibration standard and press CAL.
Cal 1 appears on the display
2
The meter endpoints according to the preselected endpoint mode after the signal has stabilized
or after pressing READ.
The calibration result is shown on the display.
3
Press Save to save the result.
— or —
18 Operating the meter
once.
once.
and
to enter the value and press READ to confirm.
and
.
and
to highlight Save.