6.3  C onductivity calibration settings
Calibration standard
- Predefined conductivity standard
The following conductivity standards are available:
International:
10 μS/cm
Chinese:
146.5 μS/cm
Japanese:
1330.00 μS/cm
- Customized conductivity standard
This option is for users who would like to use their own conductivity standard for calibration of the
con d uc t iv i t y sensor. Up to 5 temperature-dependent values (in mS/cm only) can be entered in the
table. Lowest possible special standard: 0.00005 mS/cm (0.05 μS/cm). This value corresponds
to the conductivity of pure water at 25°C, exclusively caused by the autoprotolysis of water.
When switching from a predefined standard to customized standard, you should always save the
table even if no values have changed.
- Cell constant
If the cell constant of the conductivity cell being used is accurately known, it can be entered direct-
ly in the meter.
1
Select Enter Cell Constant in the menu
2
Return to the measurement display
3
Press CAL in the measurement display
4
The user is prompted to enter the cell constant
Calibration reminder
When the calibration reminder is "On", the user is reminded to perform a new calibration after a
cer t ain user-defined interval (maximum 9999 h) has elapsed.
Press READ to save the interval and another screen appears to select calibration expiration
date.
Four different time spans can be programmed. In all four cases, a warning message appears that
the electrode should be calibrated.
Immediately
The meter is immediately blocked for measurement when the predefined interval has elapsed.
Reminder + 1h
The meter is blocked for measurement 1 hour after the predefined interval has elapsed.
Reminder + 2h
The meter is blocked for measurement 2 hours after the predefined interval has elapsed.
Continue Reading
The user can continue measuring when the predefined interval has elapsed.
28 Menus and settings
84 μS/cm
500 μS/cm
1408 μS/cm
133.00 μS/cm
1413 μS/cm
12.88 mS/cm saturated NaCl
12.85 mS/cm
111.35 mS/cm
26.6  μS/cm