4
Press Exit to reject the calibration and return to sample measurement.
The second point required for the conductivity calibration curve is permanently programmed
in the meter and is 0 S/m for a specific resistivity moving toward infinity. To ensure the most
accu r ate conductivity readings, verify the cell constant with a standard solution regularly and
recal i b rate if necessary.
4.10  S ample measurements
Place the sensor in the sample and press READ to start a measurement.
The display shows the readings of the sample.
The endpoint format blinks, indicating a measurement is in progress.
As soon as the measurement is stable, the Stability icon appears.
If the "automatic endpoint" format is selected, the measurement stops automatically as soon
as the Stability icon appears.
If the "manual endpoint" format is selected, press READ to manually stop the measurement.
If the "timed endpoint" format is selected, the measurement stops after the preset time.
4.11  D ata transfer
It is possible to transfer either all data or a user-defined set of data from the memory to a METTLER
TOLEDO printer (for example RS-P26), to a PC by using LabX direct or to a USB memory stick.
The following section describes how to proceed with the different configurations.
Data transfer from the meter to a printer
1
Connect the RS232 cable to the meter and the corresponding interface on the backside of the
print e r.
2
Select the interface "printer" in the data transfer settings menu (see "Setup: Data Transfer Set-
tings").
3
Start transfer in the data menu.
For some printers (e.g. RS-P25, RS-P26 and RS-P28), the baud rate settings will be automatically
syn c hro n ized with those of the instrument.
For other printers the settings for data transfer in the printer need to be adjusted as follows:
Baud rate: 1200
Data bits: 8
Parity: none
Stop bits: 1
Data transfer from the meter to LabX direct pH
1
Connect the instrument via USB B to the PC.
The icon appears on the display.
2
Select the interface "LabX direct" in the data transfer settings menu (see "Setup: Data Trans-
fer Set t ings").
3
Open LabX direct pH and select the correct instrument.
4
Select the item and Transfer in the data menu to start the transfer.
Operating the meter
19