Press Exit to reject the calibration and return to sample measurement.
●
The second point required for the conductivity calibration curve is permanently programmed
in the meter and is 0 S/m for a specific resistivity moving toward infinity. To ensure the most
accu r ate conductivity readings, verify the cell constant with a standard solution regularly and
recal i b rate if necessary.
4.10 S ample measurements
●
Place the sensor in the sample and press READ to start a measurement.
•
The display shows the readings of the sample.
•
The endpoint format blinks, indicating a measurement is in progress.
As soon as the measurement is stable, the Stability icon appears.
●
If the "automatic endpoint" format is selected, the measurement stops automatically as soon
as the Stability icon appears.
●
If the "manual endpoint" format is selected, press READ to manually stop the measurement.
●
If the "timed endpoint" format is selected, the measurement stops after the preset time.
4.11 D ata transfer
It is possible to transfer either all data or a user-defined set of data from the memory to a METTLER
TOLEDO printer (for example RS-P26), to a PC by using LabX direct or to a USB memory stick.
The following section describes how to proceed with the different configurations.
Data transfer from the meter to a printer
1
Connect the RS232 cable to the meter and the corresponding interface on the backside of the
print e r.
2
Select the interface "printer" in the data transfer settings menu (see "Setup: Data Transfer Set-
tings").
3
Start transfer in the data menu.
For some printers (e.g. RS-P25, RS-P26 and RS-P28), the baud rate settings will be automatically
syn c hro n ized with those of the instrument.
For other printers the settings for data transfer in the printer need to be adjusted as follows:
●
Baud rate: 1200
●
Data bits: 8
●
Parity: none
●
Stop bits: 1
Data transfer from the meter to LabX direct pH
1
Connect the instrument via USB B to the PC.
The icon appears on the display.
2
Select the interface "LabX direct" in the data transfer settings menu (see "Setup: Data Trans-
fer Set t ings").
3
Open LabX direct pH and select the correct instrument.
4
Select the item and Transfer in the data menu to start the transfer.
Operating the meter
19