The concept of the two working modes is a GLP feature that ensures that important settings and
stored data cannot be deleted cannot be unintentionally changed under routine working conditions.
The meter only allows the following functions in the routine mode:
Calibrating and measuring
Editing user, sample and sensor IDs
Editing the MTC temperature
Editing data transfer settings
Editing system-settings (PIN-protected)
Running the instrument self-test
Storing, viewing, printing and exporting data
Exporting settings to USB-stick
6. Screen settings
Screen Brightness
The screen brightness can be set from levels 1 to 16.
Screen Saver
The amount of time can be set which passes before the screen saver is activated:
5-99 minutes
If the meter is not operated during this time, the screen saver is activated. Press any key and the
dis p lay is activated again, independent of the function of the key.
Screen Color
The display background color blue, grey, red or green can be selected.
The display has a limited lifetime; therefore we recommend activating the screen saver or switching
off the meter when not in operation.
If an instrument login password is set, this password is required after activating the display again.
5.7  S ervice
1. Software Update
In case a newer software version is available, the user can perform a software update via USB-
stick, by adhering to the following process:
1
Make sure that the firmware is in the root directory of the USB-stick and has a name
S<xxx>v<yyy>.bin, with <xxx> being the number of the instrument type (220 for pH/Ion me-
ter and 230 for conductivity meter) and <yyy> being the version number.
2
Connect the USB-stick to the instrument
3
Select the option "software update"
A message appears that the software update ist in progress
4
When the software update is completed you need to restart the instrument for the changes to
become effective.
After the software update the instrument is switched back to factory settings. All unsaved data
will be lost and the PIN will be set back to "000000".
In case the USB-stick is removed during the update process or the power adapter is plugged
out, the instrument cannot be switched on anymore. Please contact METTLER TOLEDO service
in that case.