Data export from the meter to a USB stick
1
Insert the USB stick into the corresponding interface of the meter.
The icon appears on the display.
2
Select the item and Export to USB-Stick in the data menu to start the transfer.
The data will be in text (extension .txt) format. The instrument will create a new folder on the USB-
stick in which the name is the date in the international format, i.e. first year, then month and then
the day.
Example: when the date is 25 November 2011, the name of the folder will be: 20111125.
The data will be written as a text file with a name that consists of the time in 24h format (hr min
sec) with a prefix depending on what kind of data is exported. This prefix is M for measurement
data and C for calibration data.
Example: when calibration data are exported at 15:12:25 (3:12:25 pm) the name of the file will
be: C151225.txt
Pressing EXIT during the exporting aborts the process
4.12  T emperature compensation
We recommend the use of either a built-in or a separate temperature probe. If a temperature probe
is used, ATC and the sample temperature are displayed. If no temperature sensor is used, MTC is
dis p layed and the sample temperature should be entered manually.
In the conductivity mode, the meter uses this temperature to calculate with the entered alpha-coef-
fi c ient (or with non-linear correction) the conductivity value back to the chosen reference tem p er a -
ture.
20 Operating the meter