tions. The connection adapter has either a rotating pyramid adapter or a
four-hole connector for mounting of a socket adapter.
Required tools: 710D4 torque wrench, 636K13 Loctite® 241
>
1) Set the upper section of the adapter onto the prosthetic foot. Align the
adapter on the prosthetic foot in the middle of the scale.
2) Apply the lower section of the adapter on the opposite side.
3) Apply Loctite® to the 2 Allen head screws and install them (torque value:
12 Nm).
5.3 Installation of the outer sole
CAUTION
Incorrect use of prosthetic foot and outer sole
Risk of injury due to lack of surface grip and damage to the product
► Only use the prosthetic foot with an outer sole.
► Only use outer soles with spikes on sports floors suitable for this pur
pose.
1) Push the sole onto the prosthetic foot.
2) Secure the sole with the screws provided.
Prosthetic foot
5.4 Alignment
5.4.1 Bench Alignment
INFORMATION
The figures for the bench alignment can be found at the beginning of this
document.
INFORMATION
► For determining the overall system height of all prosthetic components
for bench alignment, add 5-10 mm to the socket-floor measurement to
compensate for the spring effect of the prosthetic foot. Use the larger
value when the patient's weight is at the upper limit of the maximum
permissible patient weight.
18 | Ottobock
1E91
1E93
Sole
2Z540
2Z541
2Z543