Information On Using A Cosmetic Cover; Bench Alignment - Ottobock 3R106-Pro Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 61
Fabricating a prosthesis that is optimally tailored to the patient demands technical knowledge of
orthopaedic technology. The alignment and adjustments may only be carried out by a prosthetist.
The alignment and adjustments are mainly completed in the following steps:

1. Bench alignment

2. Static alignment
3. Dynamic trial fitting
These steps are first carried out with a test prosthesis in order to determine the best possible
combination and positioning of the prosthetic components relative to each other. Once all set­
tings have been tailored to the patient, the definitive prosthesis is produced. The same steps are
repeated for this purpose.
In dynamic trial fitting, the patient must be given sufficient time to become familiar with the func­
tions of the prosthesis, thereby learning safe use for everyday life.

5.2 Information on Using a Cosmetic Cover

CAUTION
Use of talcum
Risk of injury, damage to the product due to lack of lubrication
Do not use talcum on the product or other prosthesis components.
► Only use the foam cover approved by the manufacturer (see section "Combination Possibilit­
ies" – see Page 21).
► To reduce friction and to eliminate noise, apply 519L5 Silicone Spray directly onto the contact
surfaces of the cosmetic foam cover.
► After fabricating the cosmetic cover, repeat the dynamic trial fitting since the cover affects the
adjustments of the prosthesis (e.g. damping).
5.3 Bench Alignment
Use 2Z11=KIT
INFORMATION: The connection area of the prosthetic knee joint can be protected against
scratches with the protective film of the 2Z11=KIT during alignment in the workshop and testing
in the trial fitting area.
► Use the protective film as illustrated in the accompanying document for the 2Z11=KIT.
► Remove the protective film before the patient leaves the fitting area.
+ = anterior placement / – = posterior placement (in relation to the alignment reference line)
Item
Grundaufbau
Required equipment:
Alignment apparatus (e.g., 743L200 L.A.S.A.R. Assembly or 743A200 PROS.A. Assembly),
743A80 50:50 Gauge, 743S12 Heel Height Measuring Apparatus
Position the prosthetic foot in the alignment apparatus in accordance with the following values:
   
Heel height:
effective heel height (x) + 5 mm
   
a–p positioning, middle of the prosthetic foot to the sagittal alignment reference line
+30 mm
Exterior foot rotation:
approx. 7°
m–l positioning, middle of the prosthetic foot to the frontal alignment reference line
0 mm medial
24
Bench alignment process
:
:

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

3r106-pro-kd3r106-pro-st

Tabla de contenido