Bar Code Scanner Problems; Error And Informational Messages - Philips SureSigns VS1 PM2200 Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Bar Code Scanner Problems

Bar Code Scanner Problems
Scanner does not read patient ID
Cause
Distance from bar code
Scanner needs to be reset

Error and Informational Messages

NIBP-Related Messages
Message
NIBP: Check for
Leaks
NIBP: Check Cuff/
Patient
NIBP: Motion Artifact
NIBP: Re-inflation
NIBP: Irregular Pulses
NIBP: Weak Pulse
NIBP: Timeout
NIBP: Over Pressure
NIBP: Check Cuff
Size
NIBP: Initializing
NIBP: Internal Failure
NIBP: Program in
Progress
10-6
Troubleshooting and Maintenance
Solution
Slowly move the scanner away from the bar code, starting at a distance of 3 inches,
moving away to a distance of 12 inches.
Verify that the bar code is not damaged, then scan the "Reset Bar Code Reader" bar
code that is attached to the scanner's mounting arm.
Description
Pressure in the cuff did not reach the
specified value within 30 seconds (10
seconds for neonatal).
Pressure dropped to 10 mmHg (1.3
kPa) without completing a
measurement.
Air was not discharged for 15 seconds
because of patient movement.
Cuff pressure did not exceed arterial
occlusion and cuff will inflate again.
Abnormal oscillometric waveform
Cannot measure NIBP because of
arrhythmia, patient movement, or the
pulse signal was too weak; a
measurement retry will occur.
Measurement took over 160 seconds
or more than 160 heartbeats measured
during cuff deflation
Cuff pressure rose above 325 mmHg
(43.3 kPa).
Neonatal cuff used in Adult mode.
Possible NIBP module problem found.
NIBP module failure.
NIBP interval program is running.
Action Required
Check cuff and hose connections for air leaks.
Check cuff position and placement. Verify no
patient movement occurred.
Check for patient movement or presence of
arrhythmia.
Automatic reinflation of cuff pressure; no
action required.
Check for patient movement or presence of
arrhythmia.
Check for patient movement or presence of
arrhythmia.
Verify that the appropriate cuff is being used
and applied correctly.
Check for extreme patient movement or
obstruction of air discharge.
Check cuff placement or kink in cuff or cuff
hose.
Verify and correct measurement mode and
cuff/hose selections.
The system is trying to verify the problem. No
user action is required.
Contact the Philips Response Center.
This message appears if you are attempting to
change from Neonate to Adult mode while an
interval program is running. Stop the interval
program, then change modes.

Hide quick links:

Publicidad

Tabla de contenido
loading

Tabla de contenido