Temperature Problems - Philips SureSigns VS1 PM2200 Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Temperature Problems

Temperature Problems
No temperature measurement
Cause
Incorrect model setting
Disconnected probe or cable
Probe out of well on power-
up
Defective probe
"Temp - Internal Error" shown
Cause
Defective probe
Unreliable temperature reading
Cause
If any of the following problems occur, you may also notice long temperature measurements times. In addition,
the temperature pinwheel, which indicates tissue contact, may stop moving.
Patient's mouth was open
Improper probe placement
LEDs do not change from "888" to "- - - "
Cause
Incorrect model setting
10-4
Troubleshooting and Maintenance
Solution
Ask your Biomed to reset the model number to the correct value.
Connect probe or replace it if it is defective.
Insert probe into probe well and try the measurement again.
If the "Temp-Probe Disconnected" message appears, the probe is probably
defective. Replace the probe. Insert the new probe into, out of, and back into the
probe well to reset the message.
Solution
If this message appears, the probe is probably defective. Replace the probe. Insert
the new probe into, out of, and back into the probe well to reset the message.
Solution
Ask patient to keep mouth closed during measurement.
Check the placement of the probe. In Predictive mode, the probe must be placed
directly against the sublingual artery, in the back, center of the tongue.
Solution
Ask your Biomed to reset the model number to the correct value.

Hide quick links:

Publicidad

Tabla de contenido
loading

Tabla de contenido