NIBP Specifications
Do not place the monitor in direct sunlight for an extended period of time. Direct sunlight can cause
deterioration of the liquid crystal display.
NIBP Specifications
Parameter
Measurement method
Pressure accuracy
Pulse accuracy
Measurement accuracy
Measurement range
SpO
Sensor Specifications
2
Parameter
Method
SpO
display range
2
SpO
accuracy
2
Pulse Rate display range
Pulse Rate accuracy
Temperature Specifications
Parameter
Method
Probe types
Modes
Temp. Display Range
Accuracy
Display resolution
11-2
Specifications
Specification
Oscillometric
± 1%, less than ± 3 mmHg
± 2%, or ± 2 bpm
Meets or exceeds AAMI SP-10:1992
Adult/Pediatric Mode
Systolic
60 – 250 mmHg
MAP
45 – 235 mmHg
Diastolic
40 – 200 mmHg
Pulse
40 – 200 bpm
Specification
2 wave length pulse wave type
0 – 100%
Refer to the sensor package insert for each sensor
Nellcor MAX-A disposable sensor (adult)
70 – 100% ± 2 digits
Nellcor MAX-N disposable sensor (NEO)
70 – 95% ± 2 digits
20 – 250 bpm
± 3 bpm
Specification
®
®
Alaris
Turbo*Temp
electronic predictive thermometer
Oral, Rectal
Predictive, Monitoring
Monitoring mode: 80 - 107.9
o
Predictive mode: 95 - 106
o
o
F (0.1
C)
± 0.2
(Tested with the thermometer in Monitoring mode in a calibrated water bath.)
o
o
F (0.1
C)
± 0.2
Neonate Mode
40 – 120 mmHg
30 – 100 mmHg
20 – 90 mmHg
40 – 240 bpm
o
o
F (26.6 - 42.1
C)
o
F (35 - 41.1
C)