Immunity test
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-11
Power frequency (50/
60 Hz) magnetic field
IEC 61000-4-8
U
is the AC mains voltage prior to application of the test level.
T
Connect WS Hardware Reference Manual
IEC 60601
Compliance
test level
level
<5% U
>95% voltage
T
(>95% dip
dips in U
for
T
in U
) for
0.5 cycles
T
0.5 cycle
40% U
60% voltage
T
(60% dip
dips in U
for 5
T
in U
) for 5
cycles
T
cycles
70% U
30% voltage
T
(30% dip
dips in U
for
T
in U
) for
25 cycles
T
25 cycles
<5% U
>95% voltage
T
(>95% dip
interruption
in UT) for
in U
for 5
T
5 sec
seconds
3 A/m
3 A/m
Electromagnetic Guidance and manufacturer's declaration
Electromagnetic environment guidance
A typical commercial or hospital environment mains power
source is suitable. If the user of the Connect WS requires
continued operation during power mains interruptions, it is
recommended that the Connect WS be powered from an
uninterruptible power supply.
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial or
hospital environment.