Table 1-5: Manufacturer's Declaration - Electromagnetic Immunity per IEC 60601-1-2:2014
IEC 60601 Test
Immunity Test
Level
Power
30 A/m
Frequency
Magnetic Field
IEC 61000-4-8
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
6 Vrms in ISM bands
20
Compliance
Electromagnetic Environment
Level
30 A/m
If image distortion occurs, it may be
necessary to position the FUJIFILM
SonoSite ultrasound system further
from sources of power frequency
magnetic fields or to install magnetic
shielding. The power frequency
magnetic field should be measured in
the Intended installation location to
assure that it is sufficiently low.
3 Vrms
Portable and mobile RF
communications equipment should be
used no closer to any part of the
FUJIFILM SonoSite ultrasound system
6 Vrms in ISM bands
including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended Separation Distance
d
= 1.2 √
P
Manufacturer's declaration