In the presence of magnetic fields the pump must be protected using
a ferromagnetic shield. See the appendix "Technical Information" for
detailed information.
The Turbo-V 2K-G must be connected to a primary pump (see
"Technical Information").
The Turbo-V 2K-G can be installed in any position. Fix the Turbo-V
2K-G in a stable position connecting the inlet flange of the
turbopump to a fixed counter-flange capable of withstanding a torque
of 10600 Nm around its axis.
The turbopump with ISO inlet flange must be fixed to the vacuum
chamber by means of bolts. The following table shows the necessary
number of bolts and the relevant fixing torque as recommended by
Agilent.
Tab. 1
FLANGE
ISO 250 F
The Turbo-V 2K-G cannot be fixed by means of its base. The system can only be
NOTE
fixed through its ISO 250F flange. Fixing must be done according to ISO 1609
norm. Steel bolts with a strength class at least of 500 N/mm
Turbo-V 2K-G Pumping System User Manual / 87-900-968-01 (G)
air temperature: from +5 °C to +40 °C
relative humidity: 0 - 95% (non-condensing)
FIXING DEVICE
M10 bolts
Instructions for Use
Installation
N.
FIXING TORQUE
12
22 Nm
must be used.
2
16
289/372