Siemens POINTEK CLS200 Manual De Arranque Rápido página 123

Ocultar thumbs Ver también para POINTEK CLS200:
Tabla de contenido

Publicidad

Wiring
WARNING: All field wiring must have insulation suitable for at least
250 V.
Notes:
Only qualified personnel are authorized to install and operate this equipment in
accordance with established safety practices and standards.
The Protective Earth Terminal indicated by
All wiring must be done by qualified personnel in accordance with all governing
regulations.
The equipment must be protected by a 15A fuse or circuit breaker in the building
installation.
A circuit breaker or switch in the building installation, marked as a disconnect switch,
shall be in close proximity to the equipment and within easy reach of the operator.
Use shielded cable, wire gauge 20 AWG to 14 AWG (0.5 mm
with a braided metallic shield for CE installations (or armoured cable where applicable).
Maximum working voltage between adjacent relay contacts is 250 V.
Relay contact terminals are for use with equipment which has no accessible live parts
and wiring which has insulation suitable for at least 250 V.
1.
Loosen the lid clip and remove the lid to access the connectors and electronics. (The
diagram on the next page can also be found on the underside of the lid, together with a
guide to switch function).
2.
Connect the wires to the terminals (polarity is not important).
3.
Ground the instrument according to local regulations.
4.
Tighten the gland to form a good seal.
5.
After adjusting the settings, replace the lid and secure the lid clip.
Page EN-6
Pointek CLS200/300 (Standard) – QUICK START MANUAL
must be connected to reliable ground.
2
2
to 2.0 mm
). Use a cable
7ML19985QY84

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Pointek cls300

Tabla de contenido