Test Procedures

Test the switch function:
Grasp the probe with your hand. LED L1 should turn on. LED L2 should turn on after a
slight delay, and the output relay may provide an audible click.
Test the delay:
1.
Grasp the probe with your hand. L1 should turn on (glow) immediately. The red out-
put status LED L2 should turn on after a delay.
2.
Turn the delay potentiometer P1 clockwise about 1/8 of a turn.
3.
Grasp the probe with your hand. L1 should turn on immediately, but L2 should turn
on after a longer delay than in Step 1.

Functionality Tests - CLS300

To test the basic functionality of the probe and potentiometers:
switches shown in
OFF position
O
1 2 3 4 5
F
F
/ROCKER DOWN
Preparation
1.
Ensure green power light L3 is on.
2.
Turn the delay potentiometer P1 fully clockwise (minimum delay).
3.
Set dip switches S1 to S4 OFF (full potentiometer control activated).
4.
Set switch S5 ON (high sensitivity).
Test Procedures
Test the sensitivity of the sensor:
If the yellow sensor status LED (L1) is on, turn potentiometer P2 clockwise until it
turns off.
Slowly turn the trip point potentiometer P2 counter-clockwise until the yellow
sensor status light L1 glows. Shortly afterwards, the red output status light L2 will
glow. This concludes the sensitivity test.
Page 26
Pointek CLS200/300 Standard – OPERATING INSTRUCTIONS
switch bank
P1
delay
L1
sensor
L2
status
output
(yellow)
status
(red)
P2
trip point
L3
power
(green)
7ML19985JH04