6.
Adjust the material level of the process so that the sensor is uncovered, but retains
as much buildup of material as possible on the sensor.
7.
If the yellow sensor status LED L1 is not on, turn the trip point potentiometer P2
counter-clockwise until it just turns on.
8.
Turn the trip point potentiometer P2 clockwise until LED L1 just turns off.
9.
Set DIP switch S5 to OFF (run mode).
Calibration for interface detection (Failsafe low, no delay)
1.
Ensure that the green power LED L3 is on.
2.
Set DIP switches S1 to S4 to OFF.
3.
Set DIP switch S5 to ON (set-up mode).
4.
Turn the delay potentiometer P1 fully counter-clockwise.
5.
Adjust the material level of the process so that the sensor is covered by the material
with the lowest relative dielectric constant.
6.
If the yellow sensor status LED L1 is not on, turn the trip point potentiometer P2
counter-clockwise until it just turns on.
7.
Turn the trip point potentiometer P2 clockwise until LED L1 just turns off.
8.
Set DIP switch S5 to OFF (run mode).
Delay alarm output
If you want to slow the Pointek CLS200 response, to compensate for turbulence or false
readings, set a delay interval using potentiometer P1, and set S1 and/or S2 to OFF, to
enable the delay for either alarm activation, alarm de-activation, or both.
If an immediate alarm output is critical, set the appropriate switch to ON, to disable the
delay.
The functioning of S1 and S2 depends on the alarm setting:
High alarm/
overfill protection
Low alarm/
dry run protection
To test the delay function, follow the test procedure on page 25.
Operation
After completing the setup, replace the lid and secure the lid clip. Pointek CLS200
Standard model is now ready to operate.
7ML19985JH04 Pointek CLS200/300 Standard – OPERATING INSTRUCTIONS
S1-ON disables delay of alarm de-activation (alarm OFF)
S2-ON disables delay of alarm activation (alarm ON)
S1-ON disables delay of alarm activation (alarm ON)
S2-ON disables delay of alarm de-activation (alarm OFF)