Test the delay:
1.
Turn the trip point potentiometer P2 clockwise, until the yellow sensor status LED L1
turns off.
2.
Turn the delay potentiometer P1 about 1/8 turn clockwise (delay set point).
3.
Slowly turn the trip point potentiometer P2 counter-clockwise until the yellow
sensor status light L1 glows. After an appropriate delay the red output status light L2
will glow. This concludes the delay test.
Test the switch function:
1.
Turn the delay potentiometer P1 fully counter-clockwise.
2.
Turn the trip point potentiometer P2 clockwise until the yellow sensor status LED L1
is off.
3.
Slowly turn the trip point potentiometer P2 counter-clockwise until the yellow
sensor status light glows.
4.
Slowly turn the trip point potentiometer P2 clockwise until the yellow sensor status
light just stops glowing. Grasp the probe with your hand. The yellow sensor status
light L1 will glow, indicating switch function. After an appropriate delay the red
output status light L2 will glow. This concludes the switch function test.