Siemens POINTEK CLS200 Manual De Arranque Rápido página 105

Ocultar thumbs Ver también para POINTEK CLS200:
Tabla de contenido

Publicidad

2.
Unscrew the probe sleeve from the lower assembly using two 17 mm (0.67")
wrenches across the flat surfaces, as shown below.
flat
surface
lower
wrenc
3.
Remove the heat shrink insulation covering the solder connections.
steel spacer
tension block
cable core
4.
Unsolder the connections.
: Do not cut the connections to the probe leads, as this can render them too short
Note
to work with later.
5.
Remove the tension block, and save it for re-use in step 7.
6.
Calculate the excess cable, then add back an allowance of 75 mm (3") for making the
connections:
For example, 1000 mm
7.
Cut off the excess cable.
Page 76
Pointek CLS200/300 Standard – OPERATING INSTRUCTIONS
a. Place two wrenches on the flat surfaces of the
probe as shown: hold the probe sleeve still, and turn the
lower wrench counter-clockwise to loosen the probe
lower assembly.
b. Remove the lower assembly by turning the threaded
electrode end counter-clockwise: this exposes the
three leads, the tension block, and the steel spacer.
flat
surface
= excess
less 75 mm
= allowance for connections
925 mm
= excess cable to be removed
lead
9.5 mm (0.37") heat
shrink insulation
7ML19985JH04

Publicidad