Quick Start Wizard via the handheld programmer

1. Quick Start

1.1. Quick Start Wiz
a)
Point the programmer at the display from a maximum distance of 300 mm (1 ft), then
press RIGHT arrow
b)
Press RIGHT arrow
view.
c)
Press RIGHT arrow
values and move directly to the next item.
d)
To change a setting, scroll to the desired item or key in a new value.
e)
After modifying a value, press RIGHT arrow
arrow
f)
Quick Start settings take effect only after you select Finish.
Material
Selects the appropriate echo processing algorithms for the material.
Options
Response Rate
Sets the reaction speed of the device to measurement changes in the target range.
Response Rate
SLOW
MED
FAST
Use a setting just faster than the maximum filling or emptying rate (whichever is greater).
Units
Sensor measurement units.
Options
Page EN-20
SITRANS LR250 (HART) – QUICK START MANUAL
to activate PROGRAM mode and open menu level 1.
twice to navigate to menu item 1.1 and open parameter
to open Edit mode or DOWN arrow
to move to the next item.
LIQUID
LIQUID LOW DK (low dielectric liquid)
Fill Rate/Empty Rate
M, CM, MM, FT, IN
to accept it and press DOWN
0.1 m/min (0.32 ft/min)
1.0 m/min (3.28 ft/min)
10.0 m/min (32.8 ft/min)
to accept default